Medical Physics 醫學物理學
Chapter 1 Vision and Hearing
這篇文章會向您講解DSE Physics E4 Chapter 1 Vision and Hearing 的所有concept
如果您想只在一篇文章就能知道:
- 這一課的重點概念
- 對概念的深度解說
- 必背公開考試答案
這篇文章就正合您心意!

Our eyes
- Similar to a convex lens and a screen
⇒ A real image is formed on the screen so that we can see - Only visible light (400 nm ≤ λ ≤ 750 nm) can be detected
Basic structure of our eyes

- refracted by cornea
- pass through pupil, refracted by lens
- fall on retina
- signal sent to brain via optic nerve
Refraction system
- Cornea does most of eye′s focusing;
Lens is only for fine tuning Difference in refractive indices between cornea and air is the greatest.
- To catch a sharp image, ciliary muscle may contract or relax to change the shape of lens

Light sensitive cells

rods | cones |
sensitive to different | for perception |
mainly for vision | mainly for vision |
The cones are less numerous than the rods and distributed unevenly, most concentrated at the yellow spot.
Rods and cones
- Rods and cones convert light into electrical signals (nerve impulses)
⇒ signals are sent to brain
via optic nerves - The brain interprets signals and produces vision
Two special spots on retina

- Blind spot: the point where optic nerves leave retina
⇒ no light sensitive cells
⇒ cannot detect light falling on there - Yellow spot: the point where cones are most concentrated
⇒ very sensitive to light
Power of a lens
- A thicker lens usually has a shorter focal length f
- A thicker lens is more powerful in bending light
Power of a lens
Power of a lens P:reciprocal of focal length f (in metres)
$$P=\frac1f$$
Unit: dioptre (D), 1 D = 1 m−1
- The more powerful the lens, the shorter its focal length
⇒ larger degree to which it bends light rays - By ‘real is positive’ convention,
power of convex lens is positive
Compound optical system (more than one lens)
If lenses are thin and close together,
power of combination = sum of individual power

Lens formula and the eye
$$P=\frac1f=\frac1u+\frac1\nu $$
- Treat cornea and lens as a single lens of power P
- Treat v as the size of eyeball
Accommodation
- Our eye changes the shape of lens in order to focus on an object (produces a sharp image)
⇒ accommodation - When viewing near objects, ciliary muscle contracts
⇒ lens becomes thicker - When viewing distant objects, ciliary muscle relaxes
⇒ lens becomes thinner
limits of lens
far point | near point |
farthest position at which an object can be focused | closest position at which an object can be focused |
infinity | 25 cm away |
Resolving power
- Even if we can focus on an object, we may not see all the details of it
- e.g. two headlights on a car may appear as one light spot
if the car is very far away - our eyes cannot resolve two objects when their angular separation is too small

Formula of Resolving power
For our eye to resolve two objects, their minimum angular separation has to be
$$\theta_{\min}\approx\frac{1.22\lambda}{D}\quad\text{(in radians)}$$
where
λ: wavelength of light
D: diameter of pupil
Spectral response

- Spectral response with receptor absorption curves
- Sensitivities of cones and rods vary considerably with λ
- In human eyes, for light of the same intensity, green light appears brighter than others
- Percept all visible colours as three coloured lights with different intensities
- Rods are most sensitive to light at 510 nm (blue–green) and least sensitive in red zone
In a dimly lit environment,
- only rods but no cones are used
- we cannot perceive any colours but only shades of grey
Defects of vision and corrections
- In general, defects of vision are caused by eyes that are too powerful or too weak
Short sight
- A short-sighted person can only see a near object clearly
- For a distant object, its image is formed in front of retina and therefore appears blurry
Possible causes:
- eye too powerful, or eyeball too long
- To correct short sight, reduce power of the eye with a concave lens
(∵ P < 0)
Long sight
- A person with long sight can only see a distant object clearly
- For a near object, its image is formed behind retina and therefore appears blurry
Possible causes:
- eye too weak (not powerful enough), or eyeball too short
- To correct long sight,
increase power of the
eye with a convex lens
(∵ P > 0)
Old sight
When a person gets older, lenses in the eyes lose elasticity
⇒ Lenses may not be able to change their shapes well enough to accommodate to object
For a person with both old sight and short sight
- to view a near object, ↑ Peye
⇒ convex lenses (for old sight) - to view a distant object, ↓ Peye
⇒ concave lenses (for short sight)
∴ glasses with bifocal lenses
Our ears
- Convert incoming sound waves into electrical signals
- Can detect frequencies between 20 and 20 000 Hz
Basic structure of our ears


Hearing mechanism
When eardrum is struck by some sound waves,
- it vibrates at same frequency as the waves, causing ear bones to move
- vibration is transmitted to inner ear via oval window
Pressure amplification
Pressure is amplified in two ways:
- Ear bones amplify force by
≈ 1.3 times - Area of eardrum ≈ 17 times area of oval window
Conversion to electrical signals
- Vibrations from middle ear are transmitted to cochlea
Conversion to electrical signals
- Vibrations from middle ear are transmitted to cochlea
Look at cochlea
- coiled tube consisting of three channels filled with liquids
- Upper and lower channels are connected at apex
- Vibrations enter cochlea via oval window, travel through upper to lower channel, and finally leave via round window
When cochlea is uncoiled
- non-uniform basilar membrane
between upper and lower channels - When a vibration is transmitted into cochlea, some parts of membrane vibrate more than others ⇒ resonance
Mode of resonance
- Basilar membrane near base vibrates more at high f
- Basilar membrane near apex vibrates more at low f
Mode of resonance
- Basilar membrane near base vibrates more at high f
- Basilar membrane near apex vibrates more at low f
Sensitivity of our ears
for sound waves from 60 to 1000 Hz, two sounds with Δf ≈ 2-3 Hz can be distinguished
- Cells on basilar membrane are connected to auditory nerves
- When cells are stimulated, they send signals to the brain
⇒ produces sense of hearing
Sound intensity level: Intensity
- How much energy passes through a unit area per unit time
- unit: W m−2
A sound becomes more intense if
- same amount of energy passes through a smaller area
- energy is transferred more rapidly
- more energy is transferred through same size of area
Decibel
- Our ears respond to a wide range of sound intensities,
from 10−12 to 102 W m−2 - To compare two sounds of different intensities IA and IB,
use a logarithmic scale:
$$\log_{10}I_A-\log_{10}I_B$$
- Intensity of softest sound we can just barely hear is
I0 ≈ 10−12 W m−2
$$L=10\cdot\log_{10}\left(\frac{I_A}{I_B}\right)\quad\mathrm{(in~dB)}$$
Typical sound intensity levels
the softest sound we can barely hear | 0 dB | |
whisper | 20 dB | |
conversation at 0.5 m | 65 dB | |
busy street | 85 dB | causes hearing damage after long exposure |
loud rock music | 120 dB | painful to ears |
jet engine at 50 m | 140 dB | |
space shuttle engine | 200 dB | immediate permanent |
Perception of sound: Hearing range

- The intensity level of softest sound we can barely hear is about 0 dB (= 10−12 W m−2)
- Such intensity level may vary from frequency to frequency
Loudness and phon
- Sounds that give same sense of
loudness vary with frequency ⇒ Curves of equal loudness curves are higher up at both ends
⇒ ear is less sensitive to sound of low and high frequencies
- Use phon scale to measure loudness
(which depends on listener) - On phon scale, a pure note of 1000 Hz is used for reference

Noise and hearing
- Damage to middle ear, or cochlea and nerves may arise hearing loss ⇒ hearing range may shrink
- Noise is one of causes of such damage
- Depending on exposure time to noise and its intensity, hearing loss can be temporary or permanent
- e.g. exposure to noise of 85 dB for 8 hours or more
can cause hearing loss
⇒ louder sounds, more immediate damage

For age-related loss
- the curve shifts upwards
For noise-related loss
- the curve shifts upwards
- sometimes ‘crest’ is formed ⇒ hearing was damaged more seriously for a certain range of frequencies
Chapter 2 Non-ionizing Medical Imaging
這篇文章會向您展示DSE Physics E4 Chapter 2 Non-ionizing Medical Imaging 的所有concept
如果您想只在一篇文章就能知道:
- 這一課的重點概念
- 對概念的深度解說
- 必背公開考試答案
這篇文章就正合您心意!

What is medical imaging?
Medical imaging makes use of various kinds of waves to create visual representation of our body
- Some waves are ionizing while some are non-ionizing.

Image quality
Limited by
- wavelength of the waves used
- how fine the detector is
- technique used to create visual representation from signals
Invasive and non-invasive
- Invasive means having a cut or a hole* (or a risk of having them) on a patient’s body
- But even non-invasive methods may harm the human body
e.g. exposed to ionizing radiation
Ultrasound scans
Ultrasound waves are used to produce medical images
Ultrasound scans
Ultrasound waves are used to produce medical images
- Ultrasound is a kind of sound waves
- Above upper limit of human audible frequency range
i.e. > 20 kHz
Echolocation
Key principle of ultrasound imaging
- Emit ultrasound pulses
⇒ analyse echoes
⇒ locate the target
Piezoelectric effect

- A special kind of crystal is compressed/stretched
⇒ voltage is developed across its two ends
Detecting ultrasound
- Crystal changes shape when hit by ultrasound
⇒ varying voltage is developed
⇒ electrical signal is produced
Producing ultrasound
- Apply varying voltage
⇒ crystal shape is changed
⇒ ultrasound is emitted
Transducer
- A device for producing and detecting ultrasound
- Apply ac voltage ⇒ produce ultrasound
- Measure developed voltage by CRO ⇒ detect ultrasound
Travelling of ultrasound in media
- Soft tissue & organ:
high percentage of water - Bone: mainly solid
Speed
soft tissue ≈ organ < bone
Acoustic impedance
- measures how easy ultrasound waves travel through a medium
Acoustic impedance Z
$$\mathbb{Z}=\rho\mathbb{C}$$
ρ: density of medium
c: ultrasound speed in medium
- Unit: kg m−2 s−1 or Rayl
In general, higher density ⇒ higher wave speed
⇒ Z typically increases with wave speed
Intensity reflection coefficient
- shows how much waves are reflected
- Intensity reflection coefficient α
$$\alpha=\frac{I_r}{I_0}=\frac{(Z_2-Z_1)^2}{(Z_2+Z_1)^2}$$
Note:
- Larger Z difference ⇒ larger α
- α < 1

Coupling gel
- has acoustic impedance similar to skin
- avoid reflection from air–skin boundary
Attenuation
- Intensity gradually decreases when it travels in a medium
- Reason:
- Energy being absorbed by the medium
- Ultrasound waves being scattered by the medium
Attenuation
- Intensity gradually decreases when it travels in a medium
- Reason:
- Energy being absorbed by the medium
- Ultrasound waves being scattered by the medium
- Higher ultrasound frequency
⇒ higher attenuation
⇒ shorter it can travel
Pulse-echo technique

Thickness of tissue 1 is given by
$$2s=ct\quad\Rightarrow\quad s=\frac{ct}2$$
Ultrasound scans
Information from echoes
- Echo strength
stronger echoes ⇒ larger acoustic impedance difference - Receiving time
later time ⇒ from deeper boundary
A-scan
- A stands for amplitude
- Only produce 1-D graph
- A-scan measures amplitudes of echoes
⇒ plots a graph of amplitude against time
B-scan
- B stands for brightness
- Can produce 2-D image
- B-scan uses bright dots to represent amplitude
- Transducer sends out fan beam to construct 2-D image
Resolution vs penetration
Axial resolution
- resolution along direction of beam
- shorter pulse duration
⇒ better axial resolution
Lateral resolution
- resolution normal to direction of beam
- higher frequency
⇒ better lateral resolution
Choosing suitable frequency
- Higher frequency
⇒ higher resolution & higher attenuation - For structures near surface,
use high frequency (7–12 MHz) - For deeper structures,
use low frequency (3–5 MHz)
Ultrasound Advantages and limitations
Advantages
- Non-ionizing & relatively small health effect
- Can distinguish soft tissues
- Can detect movements inside body
- Can produce 3-D images
- Inexpensive & readily available
Limitations and precautions
- Cannot scan structures covered by bones & gases
- Energy is absorbed by tissues & converted to heat
(has to monitor the power transmitted carefully) - Gas bubbles may form & damage surrounding tissues if burst
(has to monitor pressure in the region carefully )
Endoscopy
Based on total internal reflection of light

Total internal reflection
Occurs when
- n1 > n2 (i.e. towards optically less dense medium) &
- θ1 > c
Optical fibre
Basic structure
A glass fibre core surrounded by a cladding of slightly lower refractive index
Working principle
- ncore > ncladding & θ1 > c
⇒ total internal reflection occurs - Light ray is guided along the fibre
Endoscope
Contains two bundles of optical fibres
- Incoherent fibre bundle
- Coherent fibre bundle
Besides sending light, endoscope can also
- pass down tools, e.g. forceps for cutting tissues
- pass down air/water for inflating/flushing organs
Incoherent fibre bundle
sends light to the organ for illumination
Coherent fibre bundle
sends reflected light back to transmit images
Image formation
Coherent fibre bundle
- relative positions of fibres are the same between two ends
- reproduces image at eyepiece
Incoherent fibre bundle
- cannot reproduce image
- used for illumination (∵ cheaper)
Resolution
- Light transmitted in each fibre mixes up
- Each fibre forms a basic unit
For fixed diameter,
more & finer fibres
⇒ higher resolution
& the fibres can bend more
Uses
- Examine hollow organs, e.g. stomach & colon
- Directly insert into organs through natural openings,
e.g. throat & rectum - Used in keyhole surgery
- Only a few small cuts are made in body
- Can pass down surgical instruments
Endoscope Advantage
- Allow doctor to inspect the inner surfaces of organs
along tubes of the body - No ionizing radiation is used
- Used in keyhole surgery to speed up recovery
(∵ only small cuts are made) - Small tissue samples can be taken out from the body
Limitations and precautions
- Requires anaesthesia
- May require fasting to empty the organs
- Field of view is relatively narrow
- Can only view the inner surface of an organ with cavity
- Invasive, though minimal
- May cause internal bleeding/allergic reaction
Chapter 3 Ionizing Medical Imaging
這篇文章會向您展示DSE Physics E4 Chapter 3 Ionizing Medical Imaging 的所有concept
如果您想只在一篇文章就能知道:
- 這一課的重點概念
- 對概念的深度解說
- 必背公開考試答案
這篇文章就正合您心意!

Ionizing radiation
- Radiation with very high energy that can ionize an atom by knocking electrons out
⇒ can damage/kill living tissues

For EM waves,
- the higher the frequency, the more energy they carry
Ionizing radiation can penetrate deep into a material
because they have
- a small size (in form of particles) or
- a short wavelength (in form of waves)
X-rays
- High-frequency EM waves
- Can blacken a photographic film ⇒ method to detect
- Characterized by
Frequency range — 1016 to 1019 Hz
Energy range — 100 eV to 100 keV
Production of X-rays
- Electron gun shoots high-speed electrons to a heavy metal target
- Electrons decelerate rapidly during the hit
- Some KE taken away from electrons is released as X-ray
Attenuation
- When a beam of X-rays travels in a medium, its intensity gradually decreases, similar to that of ultrasound.
Causes and factors
- X-ray beam will lose energy as it ionizes the medium during its travel
- X-rays are scattered when they travel in a medium
scattered = bounced off (actually absorbed and re-emitted) in random direction
X-rays are attenuated more by a medium when
- medium has a higher density
- medium is made of element with a higher atomic number
- X-rays with a lower f are more easily scattered
Linear attenuation coefficient
For an X-ray beam of initial intensity I0 ,
- after it travels in a uniform medium for a distance of x, its transmitted intensity will drop to
$$I=I_0\cdot\mathrm{e}^{-\mu x}$$
where
μ: linear attenuation coefficient (unit: cm−1)
- The larger the coefficient, the faster the X-ray beam is attenuated to a lower intensity

Half-value thickness

Consider a graph of I against x,
- I decreases by half when it travels a
half-value thickness (x½ or HVT)
⇒ blocks of x½, 2x½ & 3x½ thick can
reduce I to 1/2, 1/4 and 1/8
⇒ transmitted I becomes negligible if
the block is 10x½ thick
Consider relation between x½ and μ
- When intensity is halved, i.e. I/I0 = ½
X-ray radiographic imaging
- X-rays are attenuated to different degrees when passing through different materials
Working principle
Analogy:
- We cannot see an object behind an umbrella, as our sight is blocked
- if intense light comes from behind, a projection of object can be seen
⇒ X-ray image is a projection of structures inside our body
Making of an X-ray image
- When an image is taken, X-ray beam passes through our body
- Tissues of our body have various compositions & thicknesses:
⇒ X-rays are attenuated to different degrees
⇒ X-rays of different intensities emerge from body
⇒ Detectors (e.g. films) visualize emergent X-rays
Appearance on image & attenuation
- The more X-rays the detector is exposed to, the darker the image that is produced
- OR, the more an X-ray beam is attenuated by a human body, the lighter the image that is produced
Matching tissues to X-ray image
e.g. a chest X-ray image
⇒ provides a high contrast between bones and neighbouring soft tissues
The formation of a radiographic image is the result of the different degrees of attenuation of X-rays by different body tissues.
Artificial contrast medium
- In an X-ray image, we can clearly distinguish bones from neighbouring soft tissues, as their contrast is sharp
- For certain organs, still possible to make them temporarily opaque by using an artificial contrast medium
⇒ compound with a large μ
⇒ can be taken orally or injected into body
e.g. to contrast between stomach and nearby tissue,
⇒ a patient take in an artificial contrast medium orally (known as a barium meal)
⇒ the medium coats inner surface of stomach
⇒ stomach becomes opaque to X-rays
⇒ a clear image is produced
Advantages
- Relatively inexpensive as compared with other imaging methods, e.g. CT scan, magnetic resonance imaging (MRI)
- Simple and fast
- Good resolution of bony structures
Disadvantages and limitations
- Ionizing, has radiation hazard
- Poor contrast of soft tissue structures
- Images of body structures overlap each other
Disadvantages and limitations
- Ionizing, has radiation hazard
- Poor contrast of soft tissue structures
- Images of body structures overlap each other
- Use of artificial contrast medium can be unpleasant;
Some may be allergic to iodine compounds (a common kind of artificial contrast medium)
Computed tomographic scan
- X-ray radiographic imaging cannot image a cross section of a body
- Use computed tomography (CT) to do so
CT images
CT scanner
- Bulkier than a X-ray imaging machine
- Basically a big gantry (like a huge doughnut) and a table
When a patient receives a CT scan, he lies on a table.
- X-ray tube inside gantry rotates around patient
- Array of X-ray detectors opposite to X-ray tube
How CT images are formed
- X-ray tube emits a fan of X-ray beams towards patient at centre
- When X-ray beams pass through a slice of patient’s body, they are attenuated
- Detectors on opposite side pick up beams and determine how much X-ray beams are attenuated
- X-ray tube rotates around, repeating process
As beams pass through tissues of different composition
and thickness,
- beams are attenuated by different degrees
- a series of emergent intensities at different angles obtained
- With help of computers, transform intensities into an array of numbers (μavg of a small part of every body slice)
- Numbers are displayed as pixels of different
shades of grey on a screen
⇒ An image of cross section of body is reconstructed
- Resolution of image depends on # pixels
⇒ Common resolution: 1024 × 1024 or above
- Smaller detectors can increase amount of X-ray data obtained, and hence resolution
CT image and attenuation coefficient
- Show both soft tissues and bones
- By convention, body parts that highly attenuate X-rays will appear white
- A single CT image can only show the plane of body
- With powerful computers, images from multiple cross sections can be combined to form a cross sectional image in perpendicular plane
A CT image is a map of attenuation coefficients of body tissues. These coefficients can be displayed in a grey or colour scale by a computer.
Back projection
Consider a certain body part,
- when X-rays are directed at it, they are attenuated by different degrees
- using shades of grey from light to dark to represent attenuation from high to low, a pattern is obtained
- Repeat the process from different angles
⇒ a series of patterns obtained - When patterns are overlapped, bones are reconstructed (light coloured spots) on image ⇒ back projection
In CT, back projection is a method of reconstructing the image of a body by measuring the data of the attenuated X-rays from multiple projections.
Advantages
- Can image cross section & locate abnormality inside body
- Good contrast between various body tissues
⇒ can take images of bones, soft-tissues and blood at the same time
Disadvantages
- A patient receiving a CT scan absorbs more X-rays than X-ray imaging
- More expensive than X-ray imaging
Comparison
CT image | X-ray image | |
image | • reconstructed from multiple • more detailed structural information • show soft tissues with good contrast | • single projection of a human body (with overlapping structures) • less detailed structural information • cannot show soft tissues with good contrast |
exposure to radiation | • take a longer time | • take a shorter time |
cost | • much more expensive | • much cheaper |
Radionuclide imaging
Waves emitted from the body?
- For both ultrasound scan and X-ray imaging, waves are directed at interior of a patient’s body from exterior
- Radionuclide imaging (RNI) is in reverse
For radionuclide imaging,
- radionuclides are mixed with suitable chemicals and taken into a patient’s body
- radionuclides will accumulate at specific organs
⇒ produce images of that organ by detecting radiation emitted
Suitable radionuclides
Kinds of radiation
Which radiation should be used for medical diagnosis?
- α sources cannot be used
∵ cannot pass through a human body
∵ highly ionizing and readily damage body tissues - Choose a γ source
∵ has highest penetrating power and
least ionizing power
Decay process
- T1/2 of a suitable radionuclide should not be too long
∵ no serious health effects - T1/2 cannot be too short
∵ medical diagnosis takes time to be carried out - Radionuclide should decay into a stable product
Other factors
Also consider
- chemical properties
- pharmaceutical properties
- cost
Technetium-99m
One of most common radionuclides in imaging because
- It decays by emitting γ rays only
$$_{43}^{99\mathrm{m}}\mathrm{Tc}\longrightarrow_{43}^{99}\mathrm{Tc}+\gamma $$
- It decays into a stable nuclide
(technetium-99) - γ rays emitted have suitable
energy and can be easily detected
One of most common radionuclides in imaging because
- Its half-life is 6 hours, not too long
or too short - Non-toxic
- Can be easily attached to various
chemicals, forming suitable tracers for
imaging different organs - Relatively inexpensive
Effective half-life
When a sample of radionuclides is taken into a patient’s body, its activity drops for two reasons
- First reason: radioactive sample
decays exponentially
⇒ with physical half-life T½p
When a sample of radionuclides is taken into a patient’s body, its activity drops for two reasons
- Second reason: sample is gradually removed from body due to biological processes (e.g. urination)
⇒ Usually a logarithmic process, or decay exponentially
⇒ with biological half-life T½b
∴ Actual time for activity inside the body to be halved
should be shorter than either T½p or T½b
⇒ effective half-life T½e
$$\begin{aligned}\frac{1}{T_{1/2\text{e}}}&=\frac{1}{T_{1/2\text{p}}}+\frac{1}{T_{1/2\text{b}}}\end{aligned}$$
- Activity A inside a human body (after a radioactive sample is uptaken) will drop exponentially with time t:
$$A=A_0•\mathrm{e}^{-kt}$$
where
A0 : initial activity
k : effective decay constant, which is
$$k=\frac{\ln2}{T_{1/2\text{e}}}=\ln2\cdot\left(\frac1{T_{1/2\text{p}}}+\frac1{T_{1/2\text{b}}}\right)$$
Imaging process
Radioactive tracer
To produce an image,
- put a radioactive tracer into a patient’s body (by injection, ingestion or inhalation)
- tracer accumulate in target organ
- detect radiation emitted and produce images
A tracer usually consists of
- a radionuclide tagged with a chemical compound
- single radionuclide,
e.g. iodine-123 and xenon-133 (gas)
⇒ Tracer can accumulate in a specific organ
Gamma camera
- A tool to detect gamma rays more effectively than films
- Four main components:

Main components:
- Collimator (a lead plate which has many holes) restricts gamma rays to enter the camera at a right angle
- Scintillator converts incoming gamma rays into visible light flashes
- Photomultiplier tubes (PMT) convert visible light flashes into electrical pulses (counts)
- Positioning logic circuit collects electrical pulses and determines the positions in human body that emit γ rays
⇒ Dots are drawn on final image to represent these positions
To form an image of the organ,
γ rays are collected over a period of time (~30 to 120 s)
Radionuclide images
Image and resolution
- By gamma camera, each gamma ray is converted into a dot on screen
- An image is accumulation of dots over a certain period of time
- More dots, better image resolution
Radionuclide images
Image and resolution
- By gamma camera, each gamma ray is converted into a dot on screen
- An image is accumulation of dots over a certain period of time
- More dots, better image resolution
- Injected tracer will accumulate in
target organs - The darker the region on the image, the greater the amount of tracer that has accumulated
Regions as seen on image
- Hot spot (dark region) indicates an accumulation of tracer that is more intense than normal
- Cold spot (light region) indicates the lack of a tracer
- Both can be due to abnormalities in a body
e.g. Black dots pointed to by arrows are hot spots
- due to abnormal growth of tissues which causes an intense uptake of tracer
- black dots may be tumours
e.g. A radioactive tracer is injected into bloodstream of patient.
- Two cold spots are seen in the lung
⇒ due to lack of tracers in these two areas of lungs
⇒ may be caused by blood clots

Dynamic imaging
- A series of images can be taken over a period of time to illustrate uptake of tracer into a particular organ
- e.g. kidney scan of a patient

- Tracer has accumulated in right kidney but not left kidney

Recording change in terms of image and counts per second
- From curve: how the activity changes with time
- From images: right kidney may have kidney stones which affect its normal function
Advantages
- Can evaluate function of organs, also monitor function of an organ following a medical treatment
- Can detect disease early (∵ pathological changes often occur before structural damages)
- Can detect disease efficiently
e.g. bone tumours can be detected quickly even if their locations are not yet known
Disadvantages

- Poor images resolution
- Post a health risk (∵ A radioactive tracer is introduced into body)
- Costly compared with ultrasound and X-ray imaging
- Usually non-specific diagnosis
e.g. a cold spot on the image of a thyroid may be due to a tumour (solid mass of tissues) or a cyst (sac filled with fluids or gas)
RNI images and X-ray images Comparison
- RNI images and X-ray images apply different electromagnetic waves (γ rays and X-rays)
RNI image | X-ray image | |
principle | • produced from radiation emitted from a specific organ • a tracer has to be introduced into patient’s body • non-invasive | • produced by projecting X-rays through a body part (and detecting how much X-rays are transmitted through) • nothing has to be put into patient’s body • non-invasive |
strength | • can study functions of organs rather than structures • can detect diseases before structural damage occurs | • can show structural detail • cannot show functions of organs |
time | more time is needed | less time is needed (1 s) |
resolution | good | poor |
Safety precautions for ionizing radiation
Effective dose
- To measure hazard, we cannot only measure how much ionizing radiation our bodies absorb
⇒ measure overall biological effect
For overall biological effect on entire human body, consider
- Absorbed dose received
- Type of radiation to be exposed
- Tissues or organs exposed to radiation
Effective dose (measured in sievert, Sv)
- measures overall biological effects on a person resulting from exposure to ionizing radiation
effective dose = absorbed dose x radiation weighting factor x tissue weighting factor
Biological effects
example of effects | occurring time | characteristic |
skin damage | acute | deterministic |
damage to reproductive system | ||
damage to blood forming system | ||
damage to digestive system | ||
damage to central nervous system | ||
cataract | latent | |
damage to immune system | ||
cancer | stochastic | |
genetic effects |
Acute or latent
When a person receives an effective dose of < 1 Sv,
- he may suffer from radiation sickness within a short time (acute effects)
- some effects may occur on him 6 months after dose is received (latent effects)
Deterministic or stochastic
For deterministic effects
- effect will not happen unless received dose exceeds
a certain value - severity increases with received dose
- e.g. If lens of eye receives an effective dose of 5 Sv,
a cataract will result
For stochastic effects
- occur by chance
- the higher the radiation dose received, the larger the chance of having these effects
- e.g. Risk of developing cancer for radiation workers is
4% per Sv, but severity of effects is independent of dose
Biological effects due to imaging
In general, a patient receives a greater radiation dose if
- a larger part of his body is exposed to radiation
- he is exposed to radiation for a longer time
- Effective dose absorbed during RNI may vary with the tracers being used
Safety precautions
Three main principles (JOD) that govern precautions
Justification of a practice
- Justify every radiological procedure with
‘the benefit against the risk’ - Always be a net benefit
Optimization of protection
- Radiation doses should be kept As Low As Reasonably Achievable (ALARA)
- e.g. radiographer should keep patient dose to a minimum as long as exposure is sufficient to produce a good diagnostic image
Dose limitation
- e.g. annual dose limit for occupational radiation workers is 20 mSv
Minimizing dose uptake
Principle to minimize dose received by staff and patients
- Time
- Distance — keep as far away as
possible from sources
⇒ well designed procedures
and radiation areas - Shielding — around radiation sources
⇒ apply dense materials such as lead and concrete
Comparison of imaging methods
| RNI image | endoscopy | X-ray radiographic imaging | X-ray CT scan | RNI | |
|---|---|---|---|---|---|
| invasive or non-invasive | non-invasive | invasive | non-invasive | non-invasive | non-invasive |
| radiation used | ultrasound | visible light | X-rays | X-rays | γ rays |
| radiation type | non-ionizing | non-ionizing | ionizing | ionizing | ionizing |
| radiation production | piezoelectric effect | light bulbs | X-ray tube | X-ray tube | radionuclides |
| principle | reflection of ultrasound waves | reflection of visible light | transmission and attenuation of X-rays from a single direction | transmission and attenuation of X-rays from multiple directions | emission of γ rays |
| major strength | no ionizing radiation and good resolution for soft tissues | no ionizing radiation | good resolution of bony structures | good contrast between various body tissues | functional study |
| major disadvantage and limitation | cannot scan structures covered by bones or air | can only view the inner surfaces of hollow organs | poor resolution for soft tissues; structures may overlap | relatively large dose of radiation | poor resolution and diagnosis not specific |
| effective dose received | nil | nil | small | medium | medium |
| time for each image | immediate | immediate | immediate | medium | long |
| allow real-time imaging | yes | yes | yes (using fluoroscopy method) | no | no |
| uses | diagnostic & surgical | diagnostic & surgical | diagnostic & surgical | diagnostic | diagnostic |
| cost | low | medium | low | medium | high |
Physics 必修課題
- Temperature and Thermometers
- Heat Capacity
- Change of State
- Gas laws and Kinetic Theory
- Motion
- Force
- More about Force
- Work Energy and Power
- Momentum
- Projectile Motion
- Uniform Circular Motion
- Gravitation
- Wave Motion
- Reflection Refraction and Diffraction
- Interference and Stationary Waves
- Light and Sound
- Reflection of Light
- Refraction of Light
- Lenses
- Electrostatics
- Circuit and Power
- AC and Domestic Electricity
- Electromagnetism
- Electromagnetic Induction
- Radiation and Radioactivity
- Rate of Decay and Uses of Radionuclides
- Nuclear Energy